3-2 2G0947 - 31JAN08
Adjustments and Replacements
2. Prepare the METER:
a. Install the BLOCK on the PROBE 203.2 mm (8.00 in.) from the
PROBE SENSOR.
b. Clean the PROBE with ALCOHOL WIPES.
c. Set the METER to display in ° C.
3. Remove the “BACK PANEL”.
F
C
METER
BLOCK
PROBE SENSOR
PROBE
203.2 mm (8.00 in.)