3-28 2G0947 - 31JAN08
Adjustments and Replacements
CAUTION:
Hot Surface
When the IMAGER is initially de-energized, the PROCESSOR
DRUM and ROLLERS are hot. Take care when removing the
PROCESSOR DRUM.
5. Release the 2 LATCHES.
6. Open the PROCESSOR COVER.
CAUTION:
The surface of the PROCESSOR DRUM is easily damaged by
fingernails and jewelry. Handle the PROCESSOR DRUM with one
hand on the DRUM PULLEY and the other hand on the opposite end.
Do not touch the silicone surface of the PROCESSOR DRUM.
When removing the PROCESSOR DRUM take care not to damage the
CONNECTOR.
7. Remove the PROCESSOR DRUM and stand it on a flat, stable surface.
To Install: 1. Reverse the steps in the removal procedure.
Note: Whenever a new PROCESSOR DRUM is installed, the temperature
adjustment for the PROCESSOR must be done (see “PROCESSOR -
DRUM Temperature” on page 3-1).
PROCESSOR
COVER
PROCESSOR
DRUM
2 LATCHES

Table of Contents