Page #103 background image
BeneVision N Series Patient Monitor Operator’s Manual 8 - 7
The TM80 is prone to offline
occasionally.
Wireless signal interference exists. â–  Put the TM80 in the MPAN or Wi-Fi
coverage area.
â–  Check the interference source and
reduce or eliminate interference.
Wireless signals are weak.
â–  Put the TM80 in the MPAN or Wi-Fi
coverage area with strong signals
strength.
â–  Remove any metallic obstructs that
stand between the TM80 and the
monitor.
Insufficient network bandwidth or
greater network delay leads to data
transfer delay.
Contact your service personnel.
The BP10 is prone to offline
occasionally.
MPAN signal interference exists.
â–  Put the BP10 in the MPAN coverage
area.
â–  Check the interference source and
reduce or eliminate interference.
MPAN signals are weak.
â–  Put the BP10 in the MPAN coverage
area with strong signals strength.
â–  Remove any metallic obstructs that
stand between the BP10 and the
monitor.
Telemetry Low Battery,
Telemetry Battery Depleted
The battery charge is low. Replace with a known good battery for the
TM80.
Telemetry NIBP Low Battery,
Telemetry NIBP Battery
Depleted
The battery charge is low. Replace with a known good battery for the
BP10.
When the TM80 is out of the
MPAN coverage area, it
cannot be connected to the
monitor by automatically
switching to Wi-Fi
connection.
The TM80 is not connected to Wi-Fi
network or its wireless network settings
are incorrect.
Verify that the TM80’s wireless network
settings are correct and the TM80 has been
connected to Wi-Fi network.
The monitor is not connected to Wi-Fi
network or its wireless network settings
are incorrect.
Verify that the monitor’s wireless network
settings are correct and the monitor has been
connected to Wi-Fi network.
Device Not Found A satellite module rack is not connected
to the monitor properly.
Replug the satellite module rack.
The monitor is not connected to Wi-Fi
network or its wireless network settings
are incorrect.
Verify that the monitor’s wireless network
settings are correct and the monitor has been
connected to Wi-Fi network.
The wireless network does not support
multicast data transfer.
The wireless network where the monitor and
the TM80 are connected needs to support
multicast data transfer.
Problem Possible Cause Corrective Action

Table of Contents

Other manuals for Mindray BeneVision N15