BeneVision N Series Patient Monitor Operator’s Manual A - 21
A.15.9 C.O. Specifications
A.15.10 ScvO2/SvO2 Specifications
A.15.10.1 ScvO
2
/SvO
2
Specifications from Vigilance II, Vigileo, and EV1000 monitor
A.15.10.2 ScvO
2
Specifications from the ScvO
2
Module
Measurement method Thermodilution method
Measurement range C.O.:
TB:
TI:
0.1 to 20 L/min
23 to 43 °C
0 to 27 °C
Resolution C.O.:
TB, TI:
0.1 L/min
0.1 °C
Accuracy C.O.:
TB, TI:
±5% or ±0.1 L /min, whichever is greater
±0.1 °C (without sensor)
Repeatability C.O.: ±2% or ±0.1 L/min, whichever is greater
Alarm range TB: 23 to 43 °C
Alarm limit Range Step
TB High (low limit + 1) to 43 °C
(low limit + 2) to 109.4 °F
0.1 °C
0.1 °F
TB Low 23 to (high limit - 1) °C
73.4 to (high limit - 2) °F
Operating mode Interfaces with Edwards Vigilance II, Vigileo, or EV1000 monitor
Measured parameter Consistent with SvO
2
/ScvO
2
-related parameters outputted by Vigilance II,
Vigileo, and EV1000 monitor
Parameter alarm SvO
2
, ScvO
2
Signal Output
Standard Meets the requirements of IEC 60601-1 for short-circuit protection and
leakage current
Output impedance ≤1000
Isolation voltage 1500 VAC
SpO
2
Analog Signal Output
Output voltage 0 to 10V (0 to 100%)
Output voltage error ±5%
Alarm Limit Range(%) Step (%)
SvO
2
/ScvO
2
High ScvO
2
<60%: (Low limit + 5%) to 60%
ScvO
2
≥60%: (Low limit + 1%) to 99%
ScvO2<60%: 5%
ScvO2≥60%: 1%
SvO
2
/ScvO
2
Low ScvO
2
<60%: 0 to (High limit - 5%)
ScvO
2
≥60%: 60% to (Low limit - 1%)
Measured parameters Measurement range Measurement accuracy