BeneVision N Series Patient Monitor Operator’s Manual 18 - 5
â– Select Stop to stop the current measurement.Select Setup to enter the C.O. menu.
â– Select Hemo Calcs to enter the Calculations menu.
• Starting a measurement without blood temperature being stable may cause measurement failure.
• The TB alarms are inactivated during a C.O. measurement, and will be reactivated automatically
after the completion of C.O. measurement.
• Please see the Instructions for Use of thermodilution catheter to determine the Comp Const and the
volume of injectate solution.
18.7 Changing C.O. Settings
18.7.1 Setting C.O. Alarm Properties
To set the C.O. alarm properties, follow this procedure:
1. Select the C.O. numeric area to enter the C.O. Measure menu.
2. Select Setup to enter the C.O. menu.
3. Select the Alarm tab.
4. Enter the password if required.
5. Set alarm properties as desired.
18.7.2 Selecting the Primary C.O. Parameter
You can select C.O. or C.I. as the main C.O. parameter. The measurement of the primary parameter displays in
larger numerics. To do so, follow this procedure:
1. Select the C.O. parameter area to enter the C.O. Measure menu.
2. Select the Setup tab.
3. Set Primary Parameter.