15 - 2 BeneVision N Series Patient Monitor Operator’s Manual
15.3 Temperature Display (for MPM and Temp Module)
The following figure shows the Temp numeric area for temperature monitoring with the MPM module or Temp
module. Your display may be configured to look different.
The following figure shows the Temp numeric area for temperature monitoring with the monitor. Your display
may be configured to look different.
15.4 Preparing for Temperature Monitoring (for MPM and Temp Module)
To prepare temperature monitoring, follow this procedure:
1. Select an appropriate probe for your patient according to patient category and measured site.
2. Plug the probe or temperature cable to the temperature connector. If you are using a disposable probe,
connect the probe to the temperature cable.
3. Follow the probe manufacturer’s instructions to connect the probe to the patient.
15.5 Changing Temperature Settings (for MPM and Temp Module)
15.5.1 Setting the Temperature Alarm Properties (for MPM and Temp Module)
To set the temperature alarm properties, follow this procedure:
1. Select the temperature numeric area to enter the Te mp menu.
2. Select the Alarm tab.
3. Enter the password if required.
4. Set the alarm properties。
15.5.2 Selecting the Temperature Label (for MPM and Temp Module)
Select the temperature label according to the measurement site. To do so, follow this procedure:
1. Select the temperature numeric area to enter the Te mp menu.
2. Select the Setup tab.
3. Set the temperature label.
15.5.3 Displaying the Temperature Difference (for MPM and Temp Module)
To display the temperature difference between two measurement sites monitored by the same temperature
module, switch on corresponding T. To do so, follow this procedure:
1. Select the temperature numeric area to enter the Te mp menu.
2. Select the Setup tab.
3. Switch on ΔT.
(1) Temperature site (2) Alarm limits (3) Temperature value
(4) Temperature difference (
ΔT): Difference between two temperature sites. It displays only when ΔT is switched on.