BeneVision N Series Patient Monitor Operator’s Manual C - 19
C.1.16 EEG Default Settings
PIP Alarm switch On
High limit 40
Low limit 1
Priority Med
Alarm Outputs Off
MVe Alarm switch On
High limit 30.0
Low limit 2.0
Priority Med
Alarm Outputs Off
Apnea Delay 20 sec
RR Source Auto
Atmosphere Temperature 25 °C
Relative humidity 55%
Paw Scale 40 cmH
2
O
Flow Scale 60 L/min
Vol Scale 1200 ml
Speed 6.25 mm/s
Paw View PIP, PEEP, Pmean, Pplat, RR
Flow View PEF, PIF, Compl
Vol View MVe, MVi, TVe, TVi
Item Default Setting
Item Default Setting
Scale 100 V
Speed 25mm/s
Low Freq Cuf-off 0.5 Hz
High Freq Cuf-off 30 H
Notch Filter On
Select Parameter SR, SEF, MF, PPF, TP, EMG
Sensor Check Montage Montage 1
Interval 30 min
Montage Setup Montage Type Bipolar Mode
EEG Expand View Default Setting
EEG EEG Channels All
Scale 100 v
Speed 25mm/s