16 - 4 BeneVision N Series Patient Monitor Operator’s Manual
16.6 Preparing for NIBP Measurements
16.6.1 Preparing the Patient for NIBP Measurements
In normal use, perform NIBP measurement on a patient who is in the following position:
â–  Comfortably seated
â–  Legs uncrossed
â–  Feet flat on the floor
â–  Back, arm and feet supported
NOTE
• It is recommended that the patient calms down and relaxes as much as possible before performing
the measurement and that the patient do not talk during the measurement.
• It is recommended to have the patient sit quietly for several minutes before taking the
measurement.
• Other factors that have been shown to result in an overestimation of blood pressure are labored
breathing, full bladder, pain etc.
16.6.2 Placing the NIBP Cuff
To place the NIBP cuff, follow this procedure:
1. Verify that the patient category setting is correct. If not, enter the Patient Management menu to change
patient category. For more information, see 5.4.2Editing Patient Information.
2. Connect the air tubing to the NIBP connector on the MPM module.
3. Select an appropriately sized cuff for the patient, and then wrap it around the limb directly over the
patient’s skin as follows:
a Determine the patient’s limb circumference.
b Select an appropriate cuff by referring to the limb circumference marked on the cuff. The width of the
cuff should be 40% (50% for neonates) of the limb circumference, or 2/3 of the length of the upper
arm or the thigh. The inflatable part of the cuff should be long enough to encircle at least 50% to 80%
of the limb.
c Apply the cuff to the patient’s upper arm or leg and make sure the Φ marking on the cuff matches the
artery location. The cuff should fit snugly, but with enough room for two fingers to be placed
between the cuff and the patient’s arm (on adults), and loosely on neonates with little or no air
present within the cuff. Otherwise it may cause discoloration and ischemia of the extremities. Make
sure that the cuff index line falls within the range markings on the cuff.
d Middle of the cuff should be at the level of the right atrium of the heart. If it is not, you must use the
measurement correction formula to correct the measurement. For more information, see
16.8.10Correcting the NIBP Measurements.
4. Connect the cuff to the air tubing. Avoid compression or restriction of pressure tubes. Air must pass
unrestricted through the tubing.
CAUTION
• A wrong cuff size and a folded or twisted bladder can cause inaccurate measurements.
• Do not touch or apply external pressure against the cuff and air tubing during NIBP measurement.
This may cause inaccurate blood pressure values.
• Use care when placing the cuff on an extremity used for monitoring other patient parameters.

Table of Contents

Other manuals for Mindray BeneVision N15

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N15 and is the answer not in the manual?

Mindray BeneVision N15 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N15
CategoryMedical Equipment
LanguageEnglish

Related product manuals