BeneVision N Series Patient Monitor Operator’s Manual 32 - 5
1. Enter the graphic trends review page.
2. Select at the upper left corner of the review page to enter the Graphic Trends Report menu.
3. Set the graphic trends report as described in 37.6.4Setting Graphic Trends Reports.
4. Select in the Graphic Trends Report menu to print the report.
32.2.7 Events Review Page
The monitor stores events in real time, including technical alarm events, physiological alarm events, manual
events, and operational events. When an event occurs, all the measurement numerics and three event-related
NOTE
waveforms 16 seconds before and after the event are stored.
• A total loss of power has no impact on the events stored.
• Alarms are saved as events and will be maintained if the equipment is powered down. The time of
equipment power down is not recorded as an event and cannot be reviewed.
• Earlier events will be overwritten by later ones if the capacity is reached.
32.2.7.1 Entering the Events Review Page
Choose one of the following methods to enter the events review page:
■ Select the Review quick key → select the Events tab.
■ Select the Main Menu quick key → from the Review column select Events.
The Event page displays event list. Events are displayed in descending chronological order. The most recent
event is displayed at the top.The number of asterisk symbols before an event indicate alarm priorities. When a
parameter label is prefixed with the plus sign (+), it indicates that the trend data is from external devices
connected to the monitor.
Different color blocks are displayed on the left of each event to indicate different event types.
â–  Red: high priority alarm event
â–  Yellow: medium priority alarm event
â–  Cyan: low priority alarm event
â–  Green: manual event
â–  White: operation-related event
The number of currently selected events and the total number of events are displayed at the top right corner of
the event list. For example, 2/4indicates that the selected event is the second event in the filtered events and the
total number of filtered events is 4. Total indicates the total number of events. For example: Tot al : 28 means that
there are a total of 28 events.
32.2.7.2 Configuring the Filter
You can filter events by time, alarm priority, alarm category, or parameter group. To configure the filter, follow
this procedure:
1. Enter the Events page.
2. Switch on Filter.
Select Filter Setup and set the desired filter criterion.
32.2.7.3 Editing Events
To edit events, follow this procedure:
1. Enter the Events page and tick off the desired events.
2. Select ... to edit the selected events.
â—† Lock: manually lock the event. Locked events cannot be deleted.
â—† Note: enter comments for the event.

Table of Contents

Other manuals for Mindray BeneVision N15

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N15 and is the answer not in the manual?

Mindray BeneVision N15 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N15
CategoryMedical Equipment
LanguageEnglish

Related product manuals