BeneVision N Series Patient Monitor Operator’s Manual C - 1
C Default Settings
C.1 Parameters Default Settings
C.1.1 ECG, Arrhythmia, ST and QT Default Settings
C.1.1.1 ECG Default Settings
Item Default Setting
HR/PR Alarm switch On
High limit Adult: 120 bpm
Pediatric: 160 bpm
Neonate: 200 bpm
Low limit Adult: 50 bpm
Pediatric: 75 bpm
Neonate: 100 bpm
Priority Med
Alarm Outputs Off
Extreme Tachy Alarm switch On
High limit Adult: 160 bpm
Pediatric: 180 bpm
Neonate: 220 bpm
Priority High
Alarm Outputs Off
Extreme Brady Alarm switch On
Low limit Adult: 35 bpm
Pediatric: 50 bpm
Neonate: 60 bpm
Priority High
Alarm Outputs Off
Alarm Source Auto
ECG1 II
ECG2 (5-lead, 6-lead, 12-lead) V, Va, V1
Va (for 6-lead only) Va
Vb(for 6-lead only) Vb
ECG Gain ×1
Speed 25 mm/sec
Filter OR: Surgery
CCU: Diagnostic
Other departments: Monitor
High FreqCut-off (for 12-lead only) 35 Hz
Notch Filter On