EasyManuals Logo

Mindray BeneVision N15 Operator's Manual

Mindray BeneVision N15
528 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #429 background imageLoading...
Page #429 background image
BeneVision N Series Patient Monitor Operator’s Manual A - 7
A.10.5 Interface Specifications of the N17/N15/N12/N12C
A.11 Signal Outputs Specifications
AC power input 1
Network connector N17: 2, standard RJ45 connectors (one on the iView module)
N15/N12/N12C: 1, standard RJ45 connector
USB connector N15/N12/N12C: 4, USB 2.0
N17: 8, USB 2.0, 4 on the iView module
Satellite module rack (SMR)Dock
connector
1 (For N17/N15, it connects the SMR, N1 Dock, or T1 Dock. For N12/N12C, it
connects the N1 Dock or T1 Dock)
Video output connector N17: 2 (one for the iView system)
N15/N12/N12C: 1
Nurse call connector 1, standard BNC
Equipotential grounding terminal 1
Auxiliary Output
Standard Meets the requirements of IEC 60601-1 for short-circuit protection and
leakage current
ECG Analog Output
Bandwidth
(-3dB; reference frequency: 10Hz)
Diagnostic mode: 0.05 to 150 Hz
Monitor mode: 0.5 to 40 Hz
Surgical mode: 1 to 20 Hz
ST mode: 0.05 to 40 Hz
Maximum QRS delay 25 ms (in diagnostic mode, and non-paced)
Gain (reference frequency 10Hz) 1V/mV (±5%)
Pace enhancement Signal amplitude: V
oh
≥2.5V
Pulse width: 10ms±5%
Signal rising and falling time: ≤100µs
IBP Analog Output
Bandwidth (-3dB; reference
frequency:1Hz)
0 to 40 Hz
Maximum transmission delay 30 ms
Gain (reference frequency 1 Hz) 1 V/100 mmHg, ±5%
Nurse Call Signal
Amplitude High level: 3.5 to 5 V, ±5%, providing a minimum of 10 mA output current;
Low level: < 0.5 V, receiving a minimum of 5 mA input current.
Rising and falling time ≤1 ms
Defib Sync Pulse
Output impedance ≤100 ohm
Maximum time delay 35 ms (R-wave peak to leading edge of pulse)
Amplitude High level: 3.5 to 5 V, ±5%, providing a maximum of 10 mA output current;
Low level: < 0.5 V, receiving a maximum of 5 mA input current.
Pulse width 100 ms ±10%
maximum rising and falling time 1 ms
Alarm output

Table of Contents

Other manuals for Mindray BeneVision N15

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N15 and is the answer not in the manual?

Mindray BeneVision N15 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N15
CategoryMedical Equipment
LanguageEnglish

Related product manuals