8 - 2 BeneVision N Series Patient Monitor Operator’s Manual
8.3.1 Pairing Procedure
Before pairing, configure the wireless network at the TM80 and at the monitor. Ensure that the TM80 and the
monitor are in the same Wi-Fi network.
To pair the TM80 with the monitor, follow this procedure:
1. Select the Main Menu quick key from the Parameters column select Wireless Module. The message
Updating... appears, indicating that the wireless module list is being updated.
2. Select the desired TM80 under the Device Name column of the Wireless Module menu. If the desired
TM80 is not displayed, select Update List.
3. Select Add on the right of the desired TM80.
4. Select Continue Patient in Monitor or Continue Patient in Telemetry.
Continue Patient in Monitor: uses the patient information and parameters settings in the monitor.
These pieces of information and settings will be synchronized to the TM80.
Continue Patient in Telemetry: uses the patient information and parameters settings at the TM80
and CMS. These pieces of information and settings will be synchronized to the monitor.
Upon completion of pairing, the message Connection Completed appears.
WARNING
Make sure that the desired TM80 is paired with the monitor.
NOTE
When Patient Category is set to Neo, the Wireless Module option is not available in the Parameters
column.
When a TM80 is paired with a BP10, the TM80 will automatically disconnect with the BP10 once it is
paired with the monitor.
If you need to change the name of TM80, see BeneVision TM80 Telemetry Monitor Operator’s Manual
(PN: 046-007479-00).
8.3.2 System Responses after Pairing a TM80 with the Monitor
This section describes system responses at the monitor and at the TM80 after pairing.
8.3.2.1 System Responses at the Monitor after Pairing a TM80 with the Monitor
Once a TM80 is paired successfully with the monitor, the monitor responds as below:
In the Wireless Module menu, the Add button on the right of the paired TM80 changes to Remove. The
symbol displays on the left of Remove.
The labels of parameters from the TM80 are suffixed with “-T, indicating that measurement values come
from the TM80. The or symbol beside the parameter label indicates that TM80 is connected with
the monitor via MPAN or Wi-Fi. The follow figure shows an example of parameter measurements from the
TM80.
The system will automatically switch between MPAN and Wi-Fi to make the TM80 stay connected with the
monitor.

Table of Contents

Other manuals for Mindray BeneVision N15

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N15 and is the answer not in the manual?

Mindray BeneVision N15 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N15
CategoryMedical Equipment
LanguageEnglish

Related product manuals