EasyManuals Logo

Mindray BeneVision N15 Operator's Manual

Mindray BeneVision N15
528 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #470 background imageLoading...
Page #470 background image
C - 6 BeneVision N Series Patient Monitor Operator’s Manual
C.1.3 SpO
2
/SpO
2
b Default Settings
Item Default Setting
SpO
2
/SpO
2
b Alarm switch On
High limit Adult: 100%
Pediatric: 100%
Neonate: 95%
Low limit 90%
Priority Med
Alarm Outputs Off
SpO
2
/SpO
2
b
Desat
Alarm switch On
Low limit 80%
Priority High
Alarm Outputs Off
SpO
2
Alarm switch On
High limit 10%
Priority Med
Alarm Outputs Off
Satsecond (for Nellcor SpO
2
)Off
NIBP Simul Off
Sensitivity (for Mindray SpO
2
)Medium
Display PI (for Mindray SpO
2
)On
Speed 25 mm/s
PR Alarm switch On
High limit Adult: 120
Pediatric: 160
Neonate: 200
Low limit Adult: 50
Pediatric: 75
Neonate: 100
Priority Med
Alarm Outputs Off
Alarm Source Auto
PR Source Auto
QRS Volume General, OR: 2
Other departments: 0
Display PR On

Table of Contents

Other manuals for Mindray BeneVision N15

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N15 and is the answer not in the manual?

Mindray BeneVision N15 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N15
CategoryMedical Equipment
LanguageEnglish

Related product manuals