10 - 6 BeneVision N Series Patient Monitor Operator’s Manual
IBP: PA IBP-S
(mmHg)
SYS × 0.75 SYS × 0.75 SYS × 1.25 SYS × 1.25 3 to 120
IBP-D
(mmHg
Dia × 0.75 Dia × 0.75 Dia × 1.25 Dia × 1.25 3 to 120
IBP-M
(mmHg)
Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25 3 to 120
IBP: CPP CPP
(mmHg)
CPP × 0.68 + 8 (CPP-15) or
35,
(whichever is
greater)
CPP × 0.86 +
35
(CPP+15) or
95,
(whichever is
smaller)
Adult: 20 to 235
Pediatric: 25 to175
Neonate: 25 to 100
IBP: CVP/
LAP/
RAP/
UVP/P1-
P4
(Venous
pressure)
IBP-M Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25 3 to 40
CO
2
EtCO
2
(mmHg)
0 to 32:
remains the
same
0 to 32:
remains the
same
0 to 32:
remains the
same
0 to 32:
remains the
same
Same as the
measurement range
33 to 35: 29 33 to 35: 29 33 to 35: 41 33 to 35: 41 Same as the
measurement range
36 to 45:
(EtCO
2
- 6)
36 to 45:
(EtCO
2
- 6)
36 to 45:
(EtCO
2
+ 6)
36 to 45:
(EtCO
2
+ 6)
Same as the
measurement range
46 to 48: 39 46 to 48: 39 46 to 48: 51 46 to 48: 51 Same as the
measurement range
>48: remains
the same
>48: remains
the same
>48: remains
the same
>48: remains
the same
Same as the
measurement range
FiCO
2
None None Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
awRR
(rpm)
awRR × 0.5 or
6 (whichever
is greater)
(awRR - 10) or
30 (whichever
is greater)
awRR × 1.5 or
30 (whichever
is smaller)
(awRR+25) or
85 rpm
(whichever is
smaller)
Adult/pediatric: 6 to 55
Neonate: 10 to 90
AG EtCO
2
Same as the CO
2
module
FiCO
2
Same as the CO
2
module
awRR
(rmp)
awRR × 0.5 or
6 (whichever
is greater)
(awRR - 10) or
30 (whichever
is greater)
awRR × 1.5 or
30 (whichever
is smaller)
awRR+25 or
85 (whichever
is smaller)
Adult/pediatric: 6 to 55
Neonate: 10 to 90
FiAA/EtAA Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
FiO
2
/EtCO
2
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
FiN
2
O/
EtN
2
O
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
Module Parameter Lower Limit Upper Limit Auto Limit Range
Adult/
pediatric
Neonate Adult/
pediatric
Neonate

Table of Contents

Other manuals for Mindray BeneVision N15

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N15 and is the answer not in the manual?

Mindray BeneVision N15 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N15
CategoryMedical Equipment
LanguageEnglish

Related product manuals