11 - 22 BeneVision N Series Patient Monitor Operator’s Manual
â—† Absolute: you can separately set the alarm properties for each ST alarm.
â—† Relative: you can set the alarm properties for ST Single and ST Dual alarms.
4. Set ST alarm properties.
11.8.8.2 Changing Leads for ST Display
The monitor automatically selects the three most deviated leads for ST display. You can also manually select the
leads. To do so, follow this procedure:
1. Select the ST numeric area, ECG numeric area, or ECG waveform area to enter the ECG menu.
2. Select the ST tab → select the Setup tab.
3. Set ST Segment. You can select up to 3 leads.
11.8.8.3 Showing ISO Point, J Point, and ST Point Marks
In the waveform area, the ISO point, J point, and ST point mark do not display on the ST segments by default. To
show these marks, follow this procedure:
1. Select the ST numeric area, ECG numeric area, or ECG waveform area to enter the ECG menu.
2. Select the ST tab→ select the Setup tab.
3. Switch on Show Markers.