BeneVision N Series Patient Monitor Operator’s Manual A - 29
Full accuracy Gases Range (%REL)1 Accuracy (%ABS)
CO
2
0≤CO
2
≤1
1<CO
2
≤5
5<CO
2
≤7
7<CO
2
≤10
CO
2
>10
±0.1
±0.2
±0.3
±0.5
Not specified
N2O 0≤NO
2
≤20
20<CO
2
≤100
±2
±3
O
2
0≤O
2
≤25
25<CO
2
≤80
80<CO
2
≤100
±1
±2
±3
Des 0≤Dev≤1
1≤Dev≤5
5≤Dev≤10
10≤Dev≤15
15≤Dev≤18
Dev>18
±0.15
±0.2
±0.4
±0.6
±1
Not specified
Sev 0≤Sev≤1
1≤Sev≤5
5≤Sev≤8
Sev>8
±0.15
±0.2
±0.4
Not specified
Enf, Iso, Hal 0≤Enf/Iso/Hal≤1
1≤Enf/Iso/Hal≤5
Enf/Iso/Hal>5
±0.15
±0.2
Not specified
awRR accuracy 2 to 60 rpm
>60 rpm
±1 rpm
Not specified
Note
1
: The highest GAS LEVEL for a single halogenated anaesthetic gas in a gas mixture that is concealed when the
anaesthetic concentration falls is 0.15/0.3% (Full/ISO accuracy).
Accuracy drift Meet the requirement for measurement accuracy within 6 hours
Apnea alarm time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Refreshing rate ≤1 s
Rise time
(10% ~ 90%)
Gas sample flow rate 120ml/min, using a DRYLINE II neonatal watertrap and sampling
line (2.5m):
CO
2
N
2
O
Hal, Iso, Sev, Des
Enf
O
2
≤250 ms
≤250 ms
≤300 ms
≤350 ms
≤600 ms
Gas sample flow rate 200ml/min, using the adult DRYLINE II water trap and sampling
line (2.5m):
CO
2
N
2
O
O
2
Hal, Iso, Sev, Des
Enf
≤250 ms
≤250 ms
≤500 ms
≤300 ms
≤350 ms
Delay time <4 s