BeneVision N Series Patient Monitor Operator’s Manual C - 3
Arrhythmia Threshold Default Settings
C.1.1.3 ST Default Settings
Vent Rhythm CCU: On
Other departments: Off
Med Off
Pause Off Low Off
Irr. Rhythm Off Prompt Off
A-Fib Off Prompt Off
PVCs/min CCU: On
Other departments: Off
Med Off
Pauses/min CCU: On
Other departments: Off
Med Off
Item Default Setting
Adult Pediatric Neonate
Asystole Delay 5 sec 5 sec 5 sec
Tachy 120 bpm 160 bpm 200 bpm
Brady 50 bpm 75 bpm 100 bpm
Extreme Tachy 160 bpm 180 bpm 220 bpm
Extreme Brady 35 bpm 50 bpm 60 bpm
Multiform PVCs Window 15 beats 15 beats 15 beats
PVCs/min 10 10 10
Pause/min 8 8 8
Pause Threshold 2.0 sec 2.0 sec 2.0 sec
AF/Irr Rhy End Time 2 min 2 min 2 min
V-Tach Rate 130 bpm 130 bpm 160 bpm
V-Brady Rate 40 bpm 40 bpm 40 bpm
V-Tach PVCs 666
V-Brady PVCs 5 5 5
Item Alarm Switch Priority Alarm Outputs
Item Default Setting
ST Alarm Mode Absolute
ST-I, ST-II, ST-III, ST-aVR, ST-aVL, ST-aVF, ST-V1, ST-
V2, ST-V3, ST-V4, ST-V5,ST-V6, ST-Va, ST-Vb (ST
Alarm Mode set to
Absolute)
Alarm switch Off
High limit 0.2 mV
Low limit -0.2 mV
Priority Med
Alarm Outputs Off