EasyManua.ls Logo

Mindray BeneVision N15

Mindray BeneVision N15
528 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C - 10 BeneVision N Series Patient Monitor Operator’s Manual
IBP-D Alarm switch On
High limit
Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 90 mmHg
Pediatric: 70 mmHg
Neonate: 60 mmHg
PA
Adult: 16 mmHg
Pediatric and neonate: 4 mmHg
Low limit
Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 50 mmHg
Pediatric: 40 mmHg
Neonate: 20 mmHg
PA
Adult: 0 mmHg
Pediatric and neonate: -4 mmHg
Priority Med
Alarm Outputs Off
IBP-M Alarm switch On
High limit
Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 110 mmHg
Pediatric: 90 mmHg
Neonate: 70 mmHg
PA
Adult: 20 mmHg
Pediatric and neonate: 26 mmHg
CVP//pCVP/ICP/RAP/LAP/UVP/P1-P4 venous pressure
Adult: 10 mmHg
Pediatric and neonate: 4 mmHg
Low limit
Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 70 mmHg
Pediatric: 50 mmHg
Neonate: 35 mmHg
PA
Adult: 0 mmHg
Pediatric and neonate: 12 mmHg
CVP/pCVP//ICP/RAP/LAP/UVP/P1-P4 venous pressure
Adult: 0 mmHg
Pediatric and neonate: 0 mmHg
Priority Med
Alarm Outputs Off
Item Default Setting

Table of Contents

Other manuals for Mindray BeneVision N15

Related product manuals