C - 10 BeneVision N Series Patient Monitor Operator’s Manual
IBP-D Alarm switch On
High limit
â– Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 90 mmHg
Pediatric: 70 mmHg
Neonate: 60 mmHg
â– PA
Adult: 16 mmHg
Pediatric and neonate: 4 mmHg
Low limit
â– Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 50 mmHg
Pediatric: 40 mmHg
Neonate: 20 mmHg
â– PA
Adult: 0 mmHg
Pediatric and neonate: -4 mmHg
Priority Med
Alarm Outputs Off
IBP-M Alarm switch On
High limit
â– Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 110 mmHg
Pediatric: 90 mmHg
Neonate: 70 mmHg
â– PA
Adult: 20 mmHg
Pediatric and neonate: 26 mmHg
â– CVP//pCVP/ICP/RAP/LAP/UVP/P1-P4 venous pressure
Adult: 10 mmHg
Pediatric and neonate: 4 mmHg
Low limit
â– Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 70 mmHg
Pediatric: 50 mmHg
Neonate: 35 mmHg
â– PA
Adult: 0 mmHg
Pediatric and neonate: 12 mmHg
â– CVP/pCVP//ICP/RAP/LAP/UVP/P1-P4 venous pressure
Adult: 0 mmHg
Pediatric and neonate: 0 mmHg
Priority Med
Alarm Outputs Off
Item Default Setting