2 - 12 BeneVision N Series Patient Monitor Operator’s Manual
You can simultaneously use maximum three IBP modules and three Temp modules (besides the Temp and IBP of
the MPM module), and two rSO
2
modules. The other modules can only be used one at a time. Otherwise, the
monitor will issue a module conflict prompt.
For example, if a CO
2
module is already loaded and then another CO
2
module is inserted, the monitor will then
prompt module conflict. To solve the problem of module conflict, just remove a module.
IBP Supports IBP
monitoring
BIS Supports BIS
monitoring
ICG Supports ICG
monitoring, Medis ICG
CCO/SvO
2
Connects Edwards Vigilance II® or Vigileo
TM
monitor, supports CCO and SvO
2
monitoring
PiCCO Supports CCO
monitoring and other hemodynamic parameters
ScvO
2
Supports ScvO
2
monitoring
EEG Supports EEG
monitoring
NMT Supports NMT
monitoring
rSO
2
Supports rSO
2
monitoring
CO
2
Supports CO
2
monitoring. The sidestream CO
2
, module can integrate O
2
(paramagnetic)
monitoring.
RM Supports RM
monitoring
AG Supports AG
monitoring. The AG module can integrate O
2
and BIS monitoring.
Tympanic Temp
adapting module
Connects the Genius tympanic thermometer to the monitor
BeneLink Connects external devices
/ Recorder
Module label Comments