BeneVision N Series Patient Monitor Operator’s Manual C - 17
FiO
2
Alarm switch On
High limit Adult and pediatric: 100%
Neonate: 90%
Low limit 18%
Priority Med
Alarm Outputs Off
EtN
2
O Alarm switch On
High limit 55%
Low limit 0%
Priority Med
Alarm Outputs Off
FiN
2
O Alarm switch On
High limit 53%
Low limit 0%
Priority Med
Alarm Outputs Off
EtAA/FiAA Alarm switch On
High limit 30%
Low limit 0.0%
Priority Med
Alarm Outputs Off
EtHal/EtEnf/EtIso Alarm switch On
High limit 3.0%
Low limit 0.0%
Priority Med
Alarm Outputs Off
FiHal/FiEnf/FiIso Alarm switch On
High limit 2.0%
Low limit 0.0%
Priority Med
Alarm Outputs Off
EtSev Alarm switch On
High limit 6.0%
Low limit 0.0%
Priority Med
Alarm Outputs Off