5-8 TM80 Telemetry Monitor Service Manual
3. On the main menu of the BP10, select System→Maintenance→NIBP
Accuracy Test.
4. Check the manometer values and the values displayed on the BP10. Both
should be 0mmHg.
5. Raise the pressure in the rigid vessel to 50 mmHg with the balloon pump. Then,
wait for 10 seconds until the measured values become stable.
6. Compare the manometer values with the values displayed on the BP10. The
difference should be 3 mmHg. If it is greater than 3 mmHg, contact your service
personnel.
7. Raise the pressure in the rigid vessel to 200 mmHg with the balloon pump.
Then, wait for 10 seconds until the measured values become stable and repeat
step 6.
NOTE
You can use an NIBP simulator to replace the balloon pump and the
reference manometer to perform the test.
You can use an appropriate cylinder and a cuff instead of the rigid
vessel.
5.4.4.10 NIBP Leakage Test
NOTE
You should perform NIBP accuracy test and make sure the test result is
pass prior to NIBP leakage test.