7-10 TM80 Telemetry Monitor Service Manual
Item Alarm Message Possible cause Solution
System
Device Error The self-test error occurs
to the TM80 main board.
The self-test error occurs
to the parameter
module, wireless module
or Mindray PAN module.
The self-test error occurs
to the parameter
module communication
or initialization.
Restart the TM80. If
the problem persists,
contact your service
personnel.
Restoring Last
Defaults Failed
Restoring the last default
configuration is error.
Loading Defaults
Failed
Loading the default
configuration is error.
No CMS
The battery of the
TM80 is used up
and the TM80 is
powered off
automatically.
The patient has
walked out of the
Wi-Fi signal
coverage area.
The Wi-Fi antenna
Replace the
battery with a
fully charged
lithium-ion
battery or new
AA battery.
Make sure that
the patient is
within the signal
coverage area.
Access the