EasyManuals Logo

Mindray BeneVision TM80 Service Manual

Mindray BeneVision TM80
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #159 background imageLoading...
Page #159 background image
TM80 Telemetry Monitor Service Manual 7-11
Item Alarm Message Possible cause Solution
System
of the TM80 fails.
The patient is not
admitted by the
CMS.
Serious WLAN
interference and
AP failure make the
TM80 unable to
connect the
network on the
CMS.
An IP address
conflict occurs.
system
maintenance
screen and
make sure that
the RSSI of
network signal
is displayed
normally.
Admit the
patient on the
CMS or
discharge the
patient at the
TM80.
Make sure that
the network
signal interface
is normal.
Make sure the IP
address of the
TM80 is unique.
Searching for
signal
The Wi-Fi signal of
the TM80 is poor.
Wi-Fi signal
interference
occurs.
Check whether
the patient is at
the edge of
signal coverage.
It is
recommended

Table of Contents