7-14 TM80 Telemetry Monitor Service Manual
Symptom Possible cause Solution
The working channel of
AP is not configured
properly.
TM80 cant work in DFS
channels in 5G band.
Contact your service personnel.
WiFi Regulatory domain
setting is wrong
Set WiFi Regulatory domain the same
with region of the hospital, and make
sure APs WiFi Regulatory domain is
also the same with region of the
hospital,
The TM80 has a fault. Check whether other TM80 monitors
can be online. If other TM80 monitors
can be online, restart the TM80 and
make sure that the configurations are
the same. If the configurations are the
same but the TM80 cannot be online,
the TM80 needs to be returned for
repair.
EAP setting is wrong Contact your service personnel.
EAP certificate is out of
date
Contact your service personnel.
The TM80 has
been
connected to
the wireless
network but
cannot be
connected to
the Central
The TM80 is not admitted
by the Central Station
yet.
Admit the TM80 again on the CMS.
IP addresses cannot be
obtained. IP addresses in
the IP address pool of the
DHCP server are used up.
Attempt to connect to the CMS
by using another network device
and check whether IP addresses
can be obtained.
Contact your service personnel.

Table of Contents

Other manuals for Mindray BeneVision TM80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision TM80 and is the answer not in the manual?

Mindray BeneVision TM80 Specifications

General IconGeneral
BrandMindray
ModelBeneVision TM80
CategoryMonitor
LanguageEnglish

Related product manuals