8-14 TM80 Telemetry Monitor Service Manual
9. Separate the pump from the integrated gap circuit, remove the fiber tape, and
take out the valve from the integrated gas circuit.
10. Take out the buzzer from the groove on the left of rear housing.
11. Remove two M1.6x3.5 cap head screws and take down the fastening sheet
metal of the battery spring.
Two M1.6x3.5 cap head screws
Remove the fiber tape
and take out the valve
from the integrated
gas circuit
Take out the buzzer