TM80 Telemetry Monitor Service Manual A-11
NOTE
If all of the applied parts correspond to the instrument type, the
applied parts will be tied together and one reading will be taken. If
any of the applied parts differ from the instrument type, all applied
parts will be tested individually, based on the type of applied part.
This applies to Auto and Step modes only.
In Case of Failure
Check any broken of the enclosure. Replace any defective part.
Inspect wiring for bad crimps, poor connections, or damage.
Test the wall outlet; verify it is grounded and is free of other wiring
abnormalities. Notify the user or owner to correct any deviations. As a
work around, check the other outlets to see if they could be used instead.
Change another probe to confirm if the fail is caused by console.
If the leakage current measurement tests fail on a new unit and if situation
can not be corrected, submit a Safety Failure Report to document the
system problem. Remove unit from operation.
If all else fails, stop using and inform the Customer Service Engineer for
analysis and disposal.
LIMITS
For CF applied parts: 50μA
For BF applied parts: 5000 μA

Table of Contents

Other manuals for Mindray BeneVision TM80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision TM80 and is the answer not in the manual?

Mindray BeneVision TM80 Specifications

General IconGeneral
BrandMindray
ModelBeneVision TM80
CategoryMonitor
LanguageEnglish

Related product manuals