EasyManua.ls Logo

NARDA EFA-200 - Recording Spatially Assigned Measurement Sequences (MATRIX)

NARDA EFA-200
212 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EFA-200/-300 7 Measurement data memory
7-7
7.1.3 Recording spatially assigned measurement sequences
(MATRIX)
Assigning measurement values to the spatial location where the measured
value was obtained is very time-consuming when a series of measurements
is evaluated. If the measured values are stored as a matrix, subsequent
processing can be simplified with suitable computer programs.
Tip: Before recording a matrix for the first time, it is a good idea to look at
the procedure as outlined in Annex B as this will give you a better
understanding of the technical background.
Note:
Measurement series using the B-Field Unit:
If you perform a spatial matrix B field measurement using the B-Field
Unit, the measurement series will be terminated as soon as you connect
the E-Field Unit to the B-Field Unit. An audible signal will be heard.
Measurement series using the E-Field Unit:
The connected B-Field Unit serves as the control unit during recording of
a spatial matrix measurement series with the E-Field Unit. The
measurement series ends automatically if you disconnect the E-Field
Unit from the B-Field Unit.
Operating sections
Operation is divided into two sections:
“MEM MATRIX” sub-menu: Defining the spatial limits and the order of the
coordinates
Matrix measurement screen: Display of measured value and target
coordinates, and operation of memory functions.
Note: Key F4 is used to switch between the two menus as required.
Activating the MEM MATRIX function
Date and time have been set correctly..
Setting the date and time: see page 8-9.
1. Press the MEM key.
The MEM menu opens.
2. Use the UP/DOWN keys to select the MATRIX function.

Table of Contents

Related product manuals