EasyManua.ls Logo

NARDA EFA-200 - Selecting the Detection Mode

NARDA EFA-200
212 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EFA-200/-300 6 Operating modes
6-3
3. Press the ENTER key to confirm your selection.
The automatically selected measurement range is shown with “AUTO”
below it in the display field F1.
Possible faults
The over-range screen is displayed.
Figure 6-2 Over-range screen
The selected range is not suitable for the field being measured. The
measurement range has been exceeded.
Select a higher measurement range.
The selected range remains active even after the “AUTO” function has
been selected.
The “MAX” function is activated.
Switch off either the “AUTO” or the “MAX” function.
6.1.3 Selecting the detection mode
Two detection modes can be selected:
RMS: The overall RMS value is calculated from the RMS values for the
three measurement axes (isotropic measurement only).
VECT PEAK: The peak value is calculated from the peak values of the
field vector. The phases of the three measurement axes are taken into
account (isotropic measurement only).
Press key F2 to toggle between the two detection modes.
The selected detection mode is shown in display field F2.

Table of Contents

Related product manuals