EFA-200/-300 6 Operating modes
6-25
6.3.8 Saving data
In Spectrum FFT mode, you can save data as follows in the data memory:
• Saving a single measurement
Saving individual values: see page 7-2
Note: Since each stored spectrum consists of a number of values, the
memory capacity is limited to about 16 spectra.
6.3.9 Possible problems
The settings (measurement range, detection mode, display mode,
frequency range) cannot be changed.
The device is in the “Stop” state.
⇒ Press the RUN key to put the device in the “Run” state.
No measurement curve is displayed.
The selected zoom is not right for the measured value axis.
⇒ Display the measured value axis in “Full Scale” mode.
Zooming in on the measured value axis: see page 6-21
Keys F1 to F4 cannot be used.
The marker is activated.
⇒ Use the ESC or RUN key to deactivate the marker.