11 Remote control EFA-200/-300
11-22
11.4.4 Optional functions
“Alarm” function
For Field Strength, Exposure STD modes only
Write / action
commands
Read commands Range of values of p
Meaning
Explanation
Activate / read visible alarm function
C50 <p> C51 0 = OFF
1 = ON
Activate / read audible alarm function
C52 <p> C53 0 = OFF
1 = ON
Define / read alarm threshold (B-field only)
C122 <p> C123 10E-9 to 31.6E-3
0.1E-3 to 316
Units T or G as
for B-field
measurement,
Write only: <SCI>
Read only:
<SCI4D>
Define / read alarm threshold (E-field only)
C124 <p> C125 1 to 100E3 Units: V/m
Write only: <SCI>
Read only:
<SCI4D>
Define / read alarm threshold referred to limit value curve
C126 <p> C127 0.1 to 100 Units: %
Write only: <SCI>
Read only:
<SCI4D>
Table 11-14 Remote control commands, “Alarm” function