EFA-200/-300 8 Optional functions
8-5
Entering current values
1. Use the UP/DOWN keys to select the required function.
2. To activate the entry fields, press the ENTER key.
The first entry field is highlighted in black.
3. Use the UP/DOWN keys to define a value.
4. Use the NEXT key to move to the next entry field.
5. Repeat steps 3 and 4 for the remaining entry fields and for the units.
6. Press ENTER to complete entry.
The current values have been selected. To save these values within a
data set of field measurement values, the current value memory must be
activated.
Note: The current values can be set in the range from 0 to 99999.9 A.
8.2.3 Activating recording of current values
The CONF CURRENT sub-menu is activated.
The current values to be recorded have been entered.
1. Use the UP/DOWN keys to select the “STORE” function.
2. Press the ENTER key until “STORE ON” is displayed.
Note: Storage of current values is automatically deactivated when you
switch the EFA off.
8.3 Choosing a limit curve
There are currently up to six limit value curves permanently stored in the
STANDARDS menu.
If the limit value curves change due to a revision of the standards, they
will be adapted by the manufacturer. They can be re-loaded with a
software update but cannot be edited by the user.
Software update: see page 10-4
The following limit value curves are currently stored:
UVV 2H/D: BGV 11 increased limit value (1999)
UVV EXP 1: BGV 11 exposure range 1 (1999)
UVV EXP 2: BGV 11 exposure range 2 (1999)
ICNIRP OCC.: ICNRP occupational (1998)
ICNIRP PUB: ICNIRP general public (1998)
VDE Exp.1: DIN VDE 0848 (Draft 1999)