6 Operating modes EFA-200/-300
6-14
6.3.2 Selecting the measurement range
There are three ways to choose the measurement range:
• Choose a value from the selection menu
• Activate the “AUTO” function.
The device automatically uses the proper measurement range.
• Activate a limit curve
The standard chosen in the STANDARDS menu is activated when
choosing the measurement range.
Choosing a limit curve: see page 8-5
Choosing a value from a selection menu
✓ The device should be in the “RUN” state.
✓ The measurement screen should be active.
1. Press the F1 key until the selection list appears.
2. Use the UP/DOWN keys to choose the measurement range.
3. Press the ENTER key to confirm your selection.
The selected measurement range will appear in display field F1.
– or –
Activating the “AUTO” function
✓ The device should be in the “RUN” state.
✓ The measurement screen should be active.
1. Press the F1 key until the selection list appears.
2. Use the UP/DOWN keys to choose the “AUTO” function.
3. Press the ENTER key to confirm your selection.
The automatically selected measurement range will appear in display
field F1 with “AUTO” below it.
– or –
Activating a limit curve
✓ The device should be in the “RUN” state.
✓ The measurement screen should be active.
1. Press the F1 key until the selection list appears.
2. Use the UP/DOWN keys to select the name of the standard entered in
the STANDARDS menu.
Choosing a limit curve: see page 8-5