6 Operating modes EFA-200/-300
6-28
6.4.3 Selecting the measurement mode (detection mode)
Note: If the “Standard” measurement range was chosen, the detection
mode is preselected and cannot be set. The F2 key and display
field are deactivated.
You can choose among three detection modes:
ACT RMS: The RMS value of each spectral line can be measured. The
measurement result is not a function of the signal’s polarization.
AVG RMS: The spectrum is averaged over a number NAVG of
successive spectra. This is a way of reducing noise, displaying lower
measured values and choosing an averaging time. The progress is
displayed using the validity indicator.
Averaging of the resulting spectrum: see page C-2
Setting the averaging parameter NAVG: see page 8-7
VECT PEAK: The peak value of individual spectral lines can be
measured.
Depending on the signal’s polarization, this measured value lies
between ACT RMS and a value 3 dB higher.
The device should be in the “RUN” state.
To switch between detection modes, press the F2 key.
The selected detection mode appears in display field F2.
6.4.4 Selecting the display mode
You have a choice of two display modes:
LIVE: The resulting spectrum is displayed.
MAX: For each spectral line, the maximum value is determined and
displayed on the screen.
The device should be in the “RUN” state.
To switch between display modes, press the F3 key.
The selected display mode appears in display field F3.
The display value “MAX” is cleared when switching to LIVE.
Note: When switching from MAX to LIVE, averaging is restarted if AVG
RMS was chosen as the detection mode.