EFA-200/-300 12 Specifications
12-7
12.3 Exposure STD (Shaped Time Domain) mode
12.3.1 Brief description
• Measurement of the magnetic flux density or electric field strength based
on a broadband measurement with simultaneous consideration of the
limit curve corresponding to the chosen setting.
• RMS and peak values are simultaneously measured and computed.
• Measurement of the magnetic flux density or electric field strength as a
percentage. 100 % indicates the limit is reached.
12.3.2 Frequency range
12.3.3 Measurement range, dynamic range
Nominal measurement range
The fixed measurement range setting is nominally 100% of the selected limit
value curve.
Dynamic range limit
The maximum exposure that can be measured is 200%, regardless of the
signal form that is present.
Broadband (+0 /-3 dB) 5 Hz to 32 kHz