EFA-200/-300 7 Measurement data memory
7-9
6. Press ENTER to complete entry.
The value for the selected coordinate has been entered. The values for
all remaining coordinates are defined in the same way.
Tip: Limits A and B can be used to define areas, lines or points, depending
on how many coordinates are set to the same value.
Changing the limit coordinates
The defined spatial area can be increased, decreased or otherwise
changed by entering new coordinates while the measurement is in
progress. In such cases, the target coordinate in the matrix measurement
screen will be set to limit “A” automatically.
Entering the coordinates of limits “A” and “B”: see page 7-8
Selecting the coordinate ranking
The coordinate ranking determines the order in which the target coordinates
for the next measurement point will be calculated. The selection depends on
the intended direction of movement within the measurement area.
Remember to use the same convention to express the direction as you used
to assign the axes defining the coordinates.
✓ The MEM MATRIX menu is open.
1. Press F1.
The MEM SEQN menu opens showing the possible ranking as a
selection list.
2. Use the UP/DOWN keys to select a ranking.
3. Press ENTER to confirm your selection.
The selected ranking will be applied to the Matrix measurement screen.
Note: When the target coordinates are calculated, the coordinate that is
shown farthest to the right in the ranking will be incremented first. If
the ranking is changed during the course of the measurement, the
target coordinate in the Matrix measurement screen will be set to
limit “A” automatically.
Tip: These settings can be saved in the device data memory for other
applications.
Saving device settings (User Setup): see page 9-4