DMTA-20073-01EN, Rev. C, November 2015
Chapter 6
94
Figure 6-20 Removing the acoustic wheel plain flange
To restore the tire’s transparency
1. Clean and wipe dry the inside of the acoustic wheel’s tire.
For more information, see “Cleaning the Liquid Chamber” on page 85.
2. Leave the liquid chamber to dry for 48 hours with the plain flange removed.
After the tire’s transparency has been restored, the acoustic wheel can be reassembled
and put back in place on the RollerFORM.