DMTA-20073-01EN, Rev. C, November 2015
Chapter 6
74
Figure 6-3 Removing the acoustic wheel
6.2 Emptying the Acoustic Wheel’s Liquid Chamber
The liquid chamber of the RollerFORM’s acoustic wheel needs to be emptied before
changing the tire, before replacing the phased array probe, or before cleaning it of
contaminants, such as algae and mold deposits, that can form in water after a period
of time.
Seat for the acoustic wheel (probe-cable side) with two half-bushings