DMTA-20073-01EN, Rev. C, November 2015
Maintenance
75
When manipulating noncorrosive glycol while operating the RollerFORM, the safety
measures found in the product’s MSDS must be applied. For more information, see
“Safety” on page 9.
When the liquid chamber of the RollerFORM is filled with water, Olympus
recommends that the water in the chamber be changed at least once a week, and that
it be emptied when it will not be used for 48 hours or more. Olympus highly
recommends filling the liquid chamber with distilled or deionized water. Failure to
follow these instructions will lead to algae formation, which requires more rigorous
maintenance. When the chamber is filled with noncorrosive glycol, it does not need to
be emptied periodically.
To empty the acoustic wheel’s liquid chamber
1. Open the OUT valve by unscrewing it one-and-a-half turns (see Figure 6-4 on
page 76).
CAUTION