70
Figure 5-7 The Indexer Clicker location
14. Roll the RollerFORM on top of the guide line, making sure the guiding laser beam
is perfectly aligned with the drawn guide line for the duration of the scan pass.
15. Repeat steps 11 to 14, until the entire area has been inspected.
16. At the end of the last guide line on the surface being inspected, immobilize the
RollerFORM. Then, making sure that the wheel on the Mini-Wheel encoder does
not turn, lift the RollerFORM off the surface.
17. Pause the instrument that is connected to the RollerFORM.
18. Save the data acquired by the instrument.
19. On the top of the laser casing, press the red On/Off button to deactivate the laser
(see Figure 5-3 on page 66).