PAGE 5-8
1118824, VER. 00
E-72 Program Error - Data written to an illegal ad-
dress.
Reboot Replace Therapy PCA
E-73 Program Error - Jump to an illegal address. Reboot Replace Therapy PCA
E-74 Program Error - Illegal instruction detected. Reboot Replace Therapy PCA
E-75 During the computation of VqSquared, it was
determined that the value would be negative,
generating this error.
Reboot
• Reseat Blower Con-
nectors
• Replace Blower Assy
• Replace Therapy PCA
E-76 Reserved for future use. Log Only None
E-77 Reserved for future use. Log Only None
E-78 Reserved for future use. Log Only None
E-79 During spinup of the motor the drive detected
that the rotor was not spinning or wires were
broken.
Reboot
• Reseat Blower Con-
nectors
• Replace Blower Assy
• Replace Therapy PCA
E-80 Parameter settings corrupted. Vent Inop Replace Therapy PCA
E-81 Parameter settings corrupted. Vent Inop Replace Therapy PCA
E-82 Program execution error or defective EE-
PROM.
Reboot Replace Therapy PCA
E-83 Defective EEPROM. Vent Inop Replace Therapy PCA
E-84 Defective EEPROM. Vent Inop Replace Therapy PCA
E-85 Reserved for future use Log Only None
E-86 Program execution error Reboot Replace Therapy PCA
E-87 Program execution error Log Only Replace Therapy PCA
E-88 3 reboots within 24 hours. Vent Inop Replace Therapy PCA
E-89 Reserved for future use Log Only None
E-90 Software rebooted since it wa initiated by the
user.
Reboot None
E-91 Spontaneous breathing has not been detected
within the alarm time.
High Alarm None
E-92 High flow condition has been detected. High Alarm None
E-93 The measured Minute Ventilation is less than
or equal to the alarm setting.
High Alarm None
E-94 High pressure condition. High Alarm Replace Therapy PCA
E-95 Excessive high pressure condition. Vent Inop Replace Therapy PCA
E-96 Low pressure condition High Alarm Replace Therapy PCA
E-97 Reserved for engineering Log Only None
E-98 Silence Key is stuck on. Log Only • Check the silence button op-
eration.
• Replace THerapy PCA.
CODE DESCRIPTION TYPE ACTION