PAGE 5-91118824, VER. 00
E-99 Input voltage is low. Medium Alarm None
E-100 Processor is unable to communicate with Baro-
metric Sensor.
Log Only Replace Therapy PCA
E-101 Indicates that the prescription on the SD card
has been accepted.
Info Alarm None
E-102 Indicates that the prescription on the SD card
has been rejected.
Info Alarm None
E-103 Program Execution Error. Vent Inop Replace Therapy PCA
E-104 Reserved for future use Log Only None
E-105 Indicates that the SD card has become corrupt-
ed
Low Alarm Install a new SD Card
E-106 Indicates that the SD card is full of data. Low Alarm Install a new SD Card
E-107 Indicates that the SD card has been removed. Info Alarm None
E-108 Indicates that the SD card has been inserted Info Alarm None
E-109 Indicates that a SD card error occurred Low Alarm Reinsert SD card
E-110 The desired tidal volume cannot be delivered
within the limits of the IPAP Min and Max set-
tings.
High Alarm None
E-111 Reserved for future use Log Only None
E-112 Unable to open BiPAP .S on the SD card. Info Alarm Install a new SD Card
E-113 Unable to read BiPAP .S on the SD card. Info Alarm Install a new SD Card
E-114 The used tried to upgrade to an older version of
a software.
Info Alarm Install a new SD Card
E-115 The user tried to upgrade to version of software
that is anot intended for this device.
Info Alarm Install a new SD Card
E-116 BiPAP .S file on the card is corrupt. Info Alarm Install a new SD Card
E-117 Program execution error Log Only Replace Therapy PCA
E-118 Therapy queue is full. Program execution error. Log Only Replace Therapy PCA
E-119 Pulse oximeter is reading high quality data. Info Alarm None
E-120 Pulse oximeter is reading poor quality data. Info Alarm None
E-121 Ventilator has collected enough pulse oximetry
data.
Info Alarm None
E-122 No SD card present to store the collected pulse
oximetry data.
Low Alarm Insert SD Card
E-123 Reserved for future use. Log Only None
E-124 Pulse Oximetry queue is full. Program execu-
tion error.
Log Only Replace Therapy PCA
E-125 Reserved for future use. Log Only Insert Card
E-126 Program Error. Reboot Replace Therapy PCA
E-127 A Thread Stack has written into reserve Log Only None
E-128 A queue has written into reserve Log Only None
CODE DESCRIPTION TYPE ACTION