PAGE 5-10
1118824, VER. 00
E-129 Program execution error Reboot Replace Therapy PCA
E-130 Indicates that the SD card has become corrupt-
ed
Low Alarm None
E-131 Indicates that the SD card has become corrupt-
ed
Low Alarm None
E-132 Indicates that a SD card error occurred Low Alarm None
E-133 Indicates that a SD card error occurred Low Alarm None
E-134 Reserved for engineering use Log Only None
E-135 Indicates that a SD card error occurred Log Only None
E-136 Indicates that a SW upgrade was started Log Only None
E-137 Reserved for future use Log Only None
E-138 Reserved for future use Log Only None
E-139 Indicates that the SD card has become corrupt-
ed
Low Alarm None
E-140 Indicates that 60W power supply was used with
heated tube humidifier
Log Only None
E-141 Program execution error Reboot Replace Therapy PCA
E-142 Heated tube temperature is at maximum (41
deg C). Could be shorted thermistor.
Log Only Replace Therapy PCA
E-143 Reserved for future use Log Only None
E-144 Indicates that the SD card has become corrupt-
ed
Low Alarm None
E-145 Indicates that the SD card has become corrupt-
ed
Low Alarm None
E-146 Reserved for future use Log Only None
E-147 Unable to create a file in the EEPROM Log Only None
E-148 Indicates that the device has started copying
logs to SD card
Info Alarm None
E-149 Indicates that the device has finished copying
logs to SD card
Info Alarm None
E-150 Indicates that the device has started copying
logs from EEPROM
Log Only None
E-151 Indicates that SD card was prematurely re-
moved during copying of logs
Log Only None
E-152 Indicates that pressure sensor may be unreli-
able
Vent Inop Replace Therapy PCA
E-153 Reserved for future use Log Only None
E-154 Indicates that the clearing of patient data failed Info Alarm None
E-155 Indicates that the clearing of patient data is in
progress
Info Alarm None
E-156 Indicates that the clearing of patient data was
successful
Info Alarm None
CODE DESCRIPTION TYPE ACTION