8.6.2 Safeguards
<M> Settings / Unit / Safeguards
Ignore lamp group malfunction
l No: When the lamp group malfunction alarm contact is
activated, then the following alarm will be generated: 'Lamp
group malfunction'.
l Yes: As soon as this setting is set to Yes, a lamp group
malfunction will be ignored for the duration of 60 hours. Once
this duration has elapsed, this setting will revert to No
automatically.
Reset ballast errors
l Yes: A reset command will be sent to each ballast currently in
error status. This setting will then revert to No automatically.
Delay ballast alarm [hh:mm:ss]
An alarm delay is the duration that an alarm situation must
continuously be present in order for an alarm to be generated. This
delay is necessary since it may take some time to receive the status
information of each ballast.
Delay modbus devices alarm
An alarm delay is the duration that an alarm situation must
continuously be present in order for an alarm to be generated.
Delay radiation alarm [hh:mm:ss]
This sets the duration that the 'Measured radiation' of the UV lamps
must continuously remain below the 'Minimum radiation level' for an
alarm message to be generated.
Delay flow alarm [hh:mm:ss]
This sets the duration that the 'Measured total flow' must continuously
remain below the 'Minimum flow' or above the 'Maximum flow' for an
alarm message to be generated.
Delay dose flow alarm [hh:mm:ss]
This sets the duration that the 'Measured total flow' must continuously
remain below the 'Calculated dose flow' minus the 'Minimum dose
flow', or above the 'Calculated dose flow' plus the 'Maximum dose
flow', for an alarm message to be generated.
Delay I min flow alarm [hh:mm:ss]
This sets the duration that the 'Calculated flow' must continuously
remain below the 'Minimum I min flow', or the 'Calculated flow' must
continuously remain above the 'Maximum I min flow', for an alarm
message to be generated.