8.8.1 Cleaning quartz tubes / Control
<M> Settings / Cleaning quartz tubes / Control
Manual control (from Synopta)
l Start: Start cleaning the quartz tubes. If the
unit is busy, then the start command will be
queued.
l Pause: Pause cleaning of the quartz tubes if
currently in progress.
l Stop: Stop cleaning of the quartz tubes if
currently in progress. If a start command for
cleaning the quartz tubes is currently queued,
then it will be removed from the queue.
Enable automatic cleaning
If Yes is selected for this setting, then the quartz
tubes may be cleaned automatically.
l No: Cleaning can only be started manually.
l Yes: Cleaning will start automatically based
on the settings 'Rad. level clean immediately',
'Rad. level clean after disinf.' and 'Quantity for
start cleaning'.
Yes
pH setpoint
This sets the pH value that the system will aim to
achieve during the cleaning process.
2
Check whether this
value is being
achieved during the
cleaning process.
Quantity for start cleaning [m3]
This sets the amount of water in cubic metres (m
3
)
that must pass through the quartz tubes in order for
the cleaning process to start.
The quartz tubes
should be cleaned at
least once per day.
Rule of thumb: Value
= 10 x rated capacity
Rad. level clean after disinf. [mW/cm²]
If the 'Measured radiation' drops below the radiation
level set here, then the quartz tubes will be cleaned
once disinfection has finished.
To be determined.
Rad. level clean immediately [mW/cm2]
If the Measured radiation drops below the radiation
level set here, then the quartz tubes will be cleaned
immediately.
To be determined.
Circulate acid duration
This sets the duration that the quartz tubes will be
cleaned by circulating a mixture of water and acid
through the tubes.
00:15:00 h
VitaLite