<M> Settings / Drain groups / Safeguards
Minimum I min flow [l/min]
If the calculated flow rate drops below the value set here,
then the alarm message 'Calc. flow at I min too low' will be
generated.
This alarm will be generated once the time set for 'Delay I
min flow alarm' has elapsed.
This setting only applies during disinfection based on a
calibration curve.
Maximum l min flow [l/min]
This sets the maximum flow rate that may be calculated
during calibration curve control.
This setting only applies during disinfection based on a
calibration curve.
Radiation dose deviation [%]
If, as a percentage, the 'Measured radiation dose' drops
below the 'Calculated radiation dose' minus the value set
here, then the alarm message 'Radiation dose too low' will
be generated.
This alarm will be generated once the time set for 'Delay
radiation dose alarm' has elapsed.
Blending ratio deviation [%]
If the 'Measured blending ratio' (the proportion of drain
water in the total water flow) drops below the 'Calculated
blending ratio' by the percentage set here, then the alarm
message 'Blending ratio too low' will be generated.
If the 'Measured blending ratio' (the proportion of drain
water in the total water flow) rises above the ‘Calculated
blending ratio' by the percentage set here, then the alarm
message 'Blending ratio too high’ will be generated.
This alarm will be generated once the time set for 'Delay
blending ratio alarm' has elapsed.
VitaLite