2 Command System RIGOL
DS2000E Programming Guide
2-89
:CURSor:MANual:AXValue?
:CURSor:MANual:BXValue?
Syntax
:CURSor:MANual:AXValue?
:CURSor:MANual:BXValue?
Description
Query the X value at cursor A (X cursor) in manual measurement with cursor.
Query the X value at cursor B (X cursor) in manual measurement with cursor.
Explanation
For the horizontal position of cursor A, refer to the
:CURSor:MANual:CAX command. The unit is
determined by the horizontal unit currently selected (refer to the
:CURSor:MANual:TUNit command).
For the horizontal position of cursor B, refer to the
:CURSor:MANual:CBX command. The unit is
determined by the horizontal unit currently selected (refer to the
:CURSor:MANual:TUNit command).
Return Format
The query returns the X value at cursor A or cursor B in manual measurement with cursor in scientific
notation.
:CURSor:MANual:AYValue?
:CURSor:MANual:BYValue?
Syntax
:CURSor:MANual:AYValue?
:CURSor:MANual:BYValue?
Description
Query the Y value at cursor A (Y cursor) in manual measurement with cursor.
Query the Y value at cursor B (Y cursor) in manual measurement with cursor.
Explanation
For the vertical position of cursor A, refer to the
:CURSor:MANual:CAY command. The unit is
determined by the vertical unit currently selected (refer to the
:CURSor:MANual:VUNit command).
For the vertical position of cursor B, refer to the
:CURSor:MANual:CBY command. The unit is
determined by the vertical unit currently selected (refer to the
:CURSor:MANual:VUNit command).
Return Format
The query returns the Y value at cursor A or cursor B in manual measurement with cursor in scientific
notation.