Remote Control
R&S
®
RTM20xx
289User Manual 1317.4726.02 ─ 01
Parameters:
<State> ON | OFF
CHANnel<m>:COUPling <Coupling>
Selects the connection of the indicated channel signal.
Suffix: 
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Coupling> DC | DCLimit | AC | ACLimit
DC
Direct connection with 50 Ω termination.
DCLimit
Direct connection with 1 MΩ termination.
AC
Connection through DC capacitor that removes the DC offset volt-
age from the input signal.
ACLimit
Connection through DC capacitor with 1 MΩ termination. The
capacitor removes the DC offset voltage from the input signal.
*RST: 
DCLimit
CHANnel<m>:SCALe <Scale>
Sets the vertical scale for the indicated channel.
Suffix: 
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Scale> Scale value, given in Volts per division.
Range:  1,00E-003  to  10; without probe attenuation
*RST:  5,00E-003
Default unit: V/div
CHANnel<m>:RANGe <Range>
Sets the voltage range across the 10 vertical divisions of the diagram. Use the command
alternativly instead of CHANnel<m>:SCALe.
Suffix: 
<m>
.
Selects the input channel. The maximum channel number is
instrument-dependent.
Command Reference