Remote Control
R&S
®
RTM20xx
306User Manual 1317.4726.02 ─ 01
Parameters:
<Polarity> POSitive | NEGative
POSitive
Positive going pulse, the width is defined from the rising to the
falling slopes.
NEGative
Negative going pulse, the width is defined from the falling to the
rising slopes.
*RST: POSitive
TRIGger:A:WIDTh:RANGe <RangeMode>
Defines how the measured pulse width is compared with the given limit(s).
Parameters:
<RangeMode> WITHin | OUTSide | SHORter | LONGer
WITHin | OUTSide
Triggers on pulses inside or outside a range defined by time ±
delta. The time is specified with TRIGger:A:WIDTh:WIDTh, the
range around is defined with TRIGger:A:WIDTh:DELTa.
SHORter | LONGer
Triggers on pulses shorter or longer than a time set with
TRIGger:A:WIDTh:WIDTh.
*RST: LONGer
TRIGger:A:WIDTh:DELTa <Delta>
Defines a range around the width value specified using TRIGger:A:WIDTh:WIDTh.
Parameters:
<Delta>
Range ±Δt ("Variation" softkey)
Range: Depends on the defined pulse width
(TRIG:A:WIDTH:WITDH)
TRIGger:A:WIDTh:WIDTh <Time1>
For the ranges WITHin and OUTSide (defined using TRIGger:A:WIDTh:RANGe), the
<Time1> defines the center of a range which is defined by the limits ±<Delta> (set with
TRIGger:A:WIDTh:DELTa).
For the ranges SHORter and LONGer, the width defines the maximum and minimum
pulse width, respectively.
Command Reference