Remote Control
R&S
®
RTM20xx
307User Manual 1317.4726.02 ─ 01
Parameters:
<Time1> Center value, maximum value or minimum value depending on the
defined range type.
Range:  2,00E-008  to  6.87194685440
Increment:  Depends on the <Time1> value
*RST:  2,00E-008
15.2.5.4 Video/TV Trigger
TRIGger:A:TV:STANdard................................................................................................307
TRIGger:A:TV:POLarity..................................................................................................307
TRIGger:A:TV:FIELd......................................................................................................307
TRIGger:A:TV:LINE........................................................................................................308
TRIGger:A:TV:STANdard <Standard>
Selects the color television standard.
Parameters:
<Standard> PAL | NTSC | SECam | PALM | I576 | P720 | P1080 | I1080
PALM = PAL-M
I576 = SDTV 576i (PAL and SECAM)
P720 | P1080 = HDTV 720/1080p (progressive scanning)
I1080 = HDTV 1080i (interlaced scanning)
*RST:  PAL
TRIGger:A:TV:POLarity <Polarity>
Selects the polarity of the signal. Note that the sync pulse has the opposite polarity. The
edges of the sync pulses are used for triggering,
See also: "Signal" on page 47
Parameters:
<Polarity> POSitive | NEGative
POSitive
If the video modulation is positive, the sync pulses are negative.
NEGative
If the modulation is negative, sync pulses are positive.
*RST:  NEGative
TRIGger:A:TV:FIELd <Field>
Sets the trigger on the beginning of the video signal fields, or on the beginning of video
signal lines.
 
Command Reference