Remote Control
R&S
®
RTM20xx
318User Manual 1317.4726.02 ─ 01
DISPlay:PERSistence:STATe <State>
Defines whether the waveform persists on the screen or whether the screen is refreshed
continuously.
Parameters:
<State> ON | OFF
ON
The waveform persists for the time defined using DISPlay:
PERSistence:TIME.
OFF
The waveform does not persist on the screen. Only the currently
measured values are displayed at any time.
*RST: OFF
DISPlay:
PERSistence:TIME <Time>
Persistence time if persistence is active (see DISPlay:PERSistence:STATe
on page 318).
Each new data point in the diagram area remains on the screen for the duration defined
here. To set infinite persistence, use DISPlay:PERSistence:INFinite.
Parameters:
<Time> Range: 5,00E-002 to Infinite
Increment: minimum 50E-3 s, increasing increment with increas-
ing persistence time
*RST: 5,00E-002
Default unit: s
DISPlay:PERSistence:INFinite <InfPersistence>
Sets the persistence time to infinite if DISPlay:PERSistence:STATe is ON. each new
data point remains on the screen infinitely until this setting is changed or the persistence
is cleared.
Parameters:
<InfPersistence> ON | OFF
*RST: OFF
DISPlay:PERSistence:TIME:AUTO <Auto>
The optimal persistence time is determined automatically by the instrument.
Parameters:
<Auto> ON | OFF
*RST: OFF
Command Reference